XC02

Following is the CDA version of XC02, last updated 10/31/95. CDA comments have not yet been incorporated.


		



AXAF

Advanced X-Ray
Astrophysics Facility

X-ray
Calibration
Test
GSE
Requirements
Document

CDR Internal Draft-09/29/95 9:54:24 AM

                                                            
Prepared by
J.W. Arenberg
TRW
AXAF Telescope Project

1    1    SCOPE                                            1

2    APPLICABLE DOCUMENTS                                  2
     2.1  REQUIRED DOCUMENTS                               2
          2.1.1 Team Member Documents                      2
               2.1.2     Government Specifications         3

3    REQUIREMENTS                                          4
                              3.1  COORDINATE SYSTEM       4
          3.1.1     XRCF                                   4
          3.1.2     Architectural Coordinate System        4
          3.1.3     MDS                                    4
          3.1.4     ATA                                    4
          3.1.5     SCIENCE INSTRUMENT MODULE FIVE AXIS
MOUNT     4
          3.1.6     HRMA                                   5
          3.1.7     HRMA X-ray Detection System            5
     3.2  UNITS                                            6

3.2.1     Length                                           6
     3.3  TIME DESIGNATION                                 7
     3.4  SUBSYSTEM REQUIREMENTS                           8
          3.4.1     MSFC Subsystems                        8
               3.4.1.1   X-ray Calibration Facility        8
               3.4.1.2   X-ray Source System               8
                    3.4.1.2.1 Source Location              8
                    3.4.1.2.2 Knowledge of Source Location 8
               3.4.1.3   X-ray Monochrometer               8
               3.4.1.4   Motion Detection System           8
               3.4.1.5   Autocollimating Telescope Assembly8
               3.4.1.6   Master Control Computer           8
          3.4.2     TRW Subsystems                         9
               3.4.2.1   Command Telemetry Unit Emulator   9

3.4.2.1.1Functional Purpose   9
                    3.4.2.1.2 Timing Signals               9
                    3.4.2.1.2.1    Frame Synchronization   9
                    3.4.2.1.2.2    FPSI Clock              9
                    3.4.2.1.2.3    CTUE Clock Stability    9
                    3.4.2.1.2.4     Facility Time Stamp    9
                    3.4.2.1.3 Data Formats                 9
                    3.4.2.1.3.1    Number of Formats       9
                    3.4.2.1.3.2    Display of Format in Use9
                    3.4.2.1.3.3    Dissemination of Data   9
                    3.4.2.1.3.4    Data to the Master
Control Computer    9
                    3.4.2.1.3.5    Data to the SI/SIM EGSE10
                    3.4.2.1.4 Network Interface           10
                    3.4.2.1.5 Interface Protocol          10
                    3.4.2.1.6 Cross Commands              10
                    3.4.2.1.6.1    Veto of Cross Commands 10
                    3.4.2.1.6.2    Cross Command Alarm    10
                    3.4.2.1.6.3    Command Inhibit on
Detection of Cross Command         10
                    3.4.2.1.6.4    Manual Restart on Cross
Command   10
                    3.4.2.1.7 CTUE Security               10
                    3.4.2.1.7.1    Password Access        10
                    3.4.2.1.7.2    Command Acceptance     10
                    3.4.2.1.8 Proximity to RCTUs          10
          3.4.3     EKC Subsystems                        12
               3.4.3.1   HRMA Support Structure           12
                    3.4.3.1.1 Weight                      12
                    3.4.3.1.2 Decentration and Initial
Alignment Tilt Tolerance 12
                    3.4.3.1.3 Remote Tilt Capability      12
                    3.4.3.1.3.1    Remote Tilt Range-Pitch12
                    3.4.3.1.3.2    Remote Tilt Range-Yaw  12
                    3.4.3.1.3.3    Remote Tilt Accuracy   12
                    3.4.3.1.3.4    Remote Tilt Step Size  12
                    3.4.3.1.3.5    Repeatability of Motion12
                    3.4.3.1.3.5.1  Tip/Tilt               12
                    3.4.3.1.3.5.2  Despace                13
                    3.4.3.1.3.5.3  Decenter               13
                    3.4.3.1.3.5.4  Clocking               13
                    3.4.3.1.4 Rotation Angle Axis Location13
                    3.4.3.1.5 Orientation Reporting       13
                    3.4.3.1.6 Repositioning Rate          13
                    3.4.3.1.7 Stability                   13
                    3.4.3.1.7.1    Thermal Stability      13
                    3.4.3.1.7.1.1  Transverse Stability   13
                    3.4.3.1.7.1.2  Axial Stability        14
                    3.4.3.1.7.2    Vibration Induced Motion14
                    3.4.3.1.7.2.1  Transverse Stability   14
                    3.4.3.1.7.2.2  Axial Stability        14
               3.4.3.2   HRMA Forward Contamination Cover 14
                    3.4.3.2.1 Obscuration of X-ray Beam   14
                    3.4.3.2.2 Time to Open                14
               3.4.3.3   HRMA Aft Contamination Cover     14
                    3.4.3.3.1 Obscuration of X-ray Beam   14
                    3.4.3.3.2 Time to Open                15
               3.4.3.4   Purge Capability                 15
               3.4.3.5   OTG Mount                        15
                    3.4.3.5.1 OTG Position Requirements   15
                    3.4.3.5.2  OTGs in Use                15
                    3.4.3.5.3 Obscuration of X-ray Beam by
Retracted OTG  15
                    3.4.3.5.4 OTG Alignment Tolerances    15
                    3.4.3.5.4.1    OTG Stability Tolerances15
                    3.4.3.5.4.2    OTG Repeatability
Tolerances     16
                    3.4.3.5.5 Insertion/Retraction Time   16
                    3.4.3.5.6 Position Feedback           16
                    3.4.3.5.7 OTG Insertion/Retraction Speed
and Acceleration    16
               3.4.3.6   Movement of the Aft Contamination
Cover and OTGs 16
               3.4.3.7   HRMA Shutter Assembly            17
                    3.4.3.7.1  Shutter Blade Clocking     17
                    3.4.3.7.2  Shutter Blade Designation  17
                    3.4.3.7.3 Simultaneity of Shutter
Operation 17
                    3.4.3.7.4 Remote Operation            17
                    3.4.3.7.5 Time to Reconfigure         17
                    3.4.3.7.6 Obscuration of x-ray beam   17
                    3.4.3.7.7 Shutter Opacity             18
                    3.4.3.7.8 Position Feedback           18
               3.4.3.8   1-g Offloader                    18
          3.4.4          SAO Subsystems                   19
               3.4.4.1   HRMA X-ray Detector System       19
                    3.4.4.1.1 Image Plane Detectors       19
                    3.4.4.1.2 Beam Normalization Detectors19
                    3.4.4.1.2.1    BND-500                19
                    3.4.4.1.2.2    BND-H                  19
               3.4.5.1   Five Axis Mount                  20
                    3.4.5.2   Physical Requirements       20
                    3.4.5.3   Envelope                    20
                    3.4.5.4   Weight                      20
               3.4.5.5        Load Test                   20
               3.4.5.6        Failures                    20
               3.4.5.7   MDS Source Accommodation         20
                    3.4.5.7.1      Location of Source     20
                    3.4.5.7.2 Source Interface Location
Stability Relative to SI Aim Point                        21
               3.4.5.8   External Thermal Interface       21
               3.4.5.9   Internal Thermal Interface       21
               3.4.5.10  Operating Modes                  21
               3.4.5.11  Static Mode                      21
               3.4.5.12  Motion                           21
                    3.4.5.12.1     X-axis:  Range         21
          3.4.6     Initial Alignment to Facility Optical
Axis 21
               3.4.6.1   Initial Alignment in X           22
               3.4.6.2   Initial Lateral Alignment        22
               3.4.6.3   Initial Alignment of Normal to SIM
Focal Plane    22
               3.4.6.4   Initial Rotation about FOA       22
               3.4.6.5   X-axis: Mechanism Resolution     22
               3.4.6.6   X-axis: Position Sensor Resolution22
               3.4.6.7   X-axis:  Rate                    22
               3.4.6.8   Y-axis:  Range                   22
               3.4.6.9   Y-axis:  Step Size               22
               3.4.6.10  Y-axis:  Resolution              22
               3.4.6.11  Y-axis:  Rate                    22
               3.4.6.12  Z-axis:  Range                   23
               3.4.6.13  Z-axis:  Step Size               23
               3.4.6.14  Z-axis:  Resolution              23
               3.4.6.15  Z-axis:  Rate                    23
               3.4.6.16  Dither  Mode                     23
                    3.4.6.16.1     Step Size              23
                    3.4.6.16.2     Range of Capability    23
                    3.4.6.16.3     Positonal Accuracy-
Absolute  23
                    3.4.6.16.4     Postional Accuracy-
Relative to Starting Position      23
                    3.4.6.16.5                            24
                    3.4.6.16.6     Motion Control         24
                    3.4.6.16.7     Data                   24
                    3.4.6.16.7.1   Data Display           24
                    3.4.6.16.7.2   Data Transmission      24
          3.4.7     Stability                             25
               3.4.7.1   Thermal Stability                25
                    3.4.7.1.1 Transverse Stability        25
                    3.4.7.1.2 Axial Stability             25
               3.4.7.2   Vibration Induced Motion         25
                    3.4.7.2.1 Transverse Stability        25
                    3.4.7.2.2 Axial Stability             25
                3.4.8Rotation                             25
               3.4.8.1   X-axis: Rotation                 25
               3.4.8.2   Y-axis:  Range                   26
               3.4.8.3   Y-axis:  Step Size               26
               3.4.8.4   Y-axis:  Resolution              26
               3.4.8.5   Y-axis:  Rate                    26
               3.4.8.6   Z-axis:  Range                   26
               3.4.8.7   Z-axis:  Step Size               26
               3.4.8.8   Z-axis:  Resolution              26
          3.4.9     FAM Controller Software               26
               3.4.9.1   Communication                    26
               3.4.9.2   Data Recording                   26
          3.4.10    ACIS Cryogenic Interface              26
          3.4.11    Access                                27
          3.4.12    Vacuum Interface                      27
          3.4.13    HRC Subsystems                        28
               3.4.13.1  Electrical Ground Support Equipment-
HRC  28
                    3.4.13.1.1     Command Validation     28
                    3.4.13.1.2     Telemetry Decommutation28
               3.4.13.2  Verification                     28
               3.4.13.3  Command Format                   28
               3.4.13.4  SI Command Log                   28
          3.4.14    ACIS Subsystems                       29
               3.4.14.1  Electrical Ground Support Equipment-
ACIS 29
                    3.4.14.1.1     Command Validation     29
                    3.4.14.1.2     Telemetry Decommutation29
               3.4.14.2  Verification                     29
               3.4.14.3  Command Format                   29
               3.4.14.4  SI Command Log                   29
          3.4.15    LETG Subsystems                       30
          3.4.16    HETG Subsystems                       31
          3.4.17    ASC Subsystems                        32
          3.4.18    Other Systems                         33
               3.4.18.1  X-ray  Rehearsal Optic           33
                    3.4.18.1.1     Spot Size              33
                    3.4.18.1.2     Effective Area         33
                    3.4.18.1.3     f/number               33
               3.4.18.2  X-ray  Rehearsal Optic Mount     33
                    3.4.18.2.1     Postitioning           33
                    3.4.18.2.2     Stability              33

4    ALIGNMENT/STABILITY REQUIREMENTS AND ALLOCATIONS     34
     4.1  HRMA TO FACILITY OPTICAL AXIS                   34
          4.1.1     AXIAL LOCATION                        34
          4.1.2     LATERAL DECENTRATION OF HRMA NODAL
POINT:    34
          4.1.3     TILT OF THE HRMA OPTICAL AXIS:        34
          4.1.4     HRMA CLOCKING                         34
     4.2  HRMA TO OTG                                     35
          4.2.1     Static Alignment Tolerances           35
               4.2.1.1   Static Alignment in X            35
               4.2.1.2   Static Decenter                  35
               4.2.1.3   Static Alignment in Rotation About
X    35
               4.2.1.4   Static Alignment in Rotation About
Y    35
               4.2.1.5   Static Alignment in Rotation About
Z    35
          4.2.2     Stability Tolerances                  35
               4.2.2.1   Stability in X                   35
               4.2.2.2   Decenter Stability               35
               4.2.2.3   Stability in Rotation About X    35
               4.2.2.4   Stability in Rotation About Y    36
               4.2.2.5   Stability in Rotation About Z    36
          4.2.3     Repeatability Tolerances              36
               4.2.3.1   Repeatability in Despace         36
               4.2.3.2   Repeatability in Decenter        36
               4.2.3.3   Repeatability in Rotation About X36
               4.2.3.4   Repeatability in Rotation About Y36
               4.2.3.5   Repeatability in Rotation About Z36
     4.3  ENVIRONMENT                                     37
          4.3.1     Stray Light                           37
          4.3.2     Microseismic Vibration                37
          4.3.3     Vacuum                                37
          4.3.4     Thermal                               37

5         WORKMANSHIP STANDARDS                           38
     5.1  BAKEOUT                                         38
          5.1.1     Surface Cleanliness Levels            38
     5.2  TRAPPED VOLUMES                                 39

6         GLOSSARY OF TERMS                               40

7    ACRONYMS                                             41



1     1   SCOPE1    SCOPE

                                                            
It is the scope of this document to state the requirements for
the ground support equipment (GSE) needed to calibrate the
Advanced X-ray Astrophysics Facility-Imaging (AXAF-I) high
resolution mirror (HRMA) and its science complement of science
instruments (SI).  This calibration activity is to be carried out
at the George C. Marshall Space Flight Center (MSFC) in the x-ray
calibration facility (XRCF).


2    APPLICABLE DOCUMENTSPPLICABLE DOCUMENTS

2.1    REQUIRED DOCUMENTS REQUIRED DOCUMENTS

2.1.1      Team Member DocumentsTeam Member Documents

MSFC

MSFC-SPEC-1238           Thermal Vacuum Bakeout Specification for
Contamination Sensitive Hardware

MSFC-SPEC-1837           AXAF X-ray Test Calibration Facility
Requirements

MSFC-SPEC-1839           AXAF X-ray HRMA/SI Calibration
Requirements

MSFC-RQMT-2229           Scientific Requirements for AXAF-I
Calibration

FAC-EJ-4708              XRCF Specification and Drawing Package

MSFC-SPEC-2401           End Item Specification for the X-ray
Calibration Facility X-ray Source System

TRW Documents

DPD 692 XC05             XRCF GSE Interface Definition Document

DPD 692 SE28             AXAF Contamination and Control Plan

EQ16-0057                Equipment Specification for Command
Telemetry Unit Emulator

D17830                   Motion Detection System Requirements

DR XC02                   VETA-I Calibration Requirements.

TRW Drawings

E445905                  XRCF Coordinate System

E445907                  Architectural Coordinate System

E445908                  MDS Coordinate System and Sign
Convention

E445909                  HRMA Shutter Blade Designations

E445910                  Alignment Telescope Assembly Coordinate
System

E445900                  XRCF Data System

2.1.2     Government Specificationsovernment Specifications

Federal

FED-STD-209              Clean Room and Work Station
Requirements, Controlled Environment

Military

MIL-STD-1246             Product Cleanliness Levels and
Contamination Control Program
3
REQUIREMENTSEQUIREMENTS

3.1  COORDINATE SYSTEMOORDINATE SYSTEM

There are many coordinate systems used in the execution of x-ray
test activities.  It is important that the relationships among
the various systems be well understood in order to facilitate
proper and accurate communication among team members.

3.1.1     XRCFRCF

The XRCF coordinate system is defined in TRW drawing (DWG)
E445905.

3.1.2     Architectural Coordinate Systemrchitectural Coordinate
System

The architectural coordinate system (ACS) is based upon surveying
datum used in the design of the XRCF.  The ACS is based on
stations (STA) located 100 feet apart on the guide tube (GT)
centerline.  The west end of the GT is denoted as STA 27+01.5 and
the station number decreases to the east.  This coordinate system
is shown in DWG E445907 and is used primarily to assign positions
of GT features.

3.1.3     MDSDS

The MDS coordinate system is used to reckon the relative motion
data reported by the MDS.  The axes of the MDS coordinate system
are parallel to but displaced from the XRCF coordinate system
defined above.  The MDS X axis is parallel to the XRCF X axis and
is defined by the line in the XRCF coordinate system Y=-4.76
inches, Z=-2.75 inches.  The MDS Y and Z axes are parallel to the
respective XRCF axis.  The MDS origin is at the OPS, located
proximate to the focal plane of the HRMA.  The axial displacement
of the origin is irrelevant since the MDS can only reckon
relative motion in the MDS or XRCF (Y, Z) plane.

The MDS coordinate system and sign convention are given in DWG
E445908.

3.1.4     ATATA

The ATA coordinate system is defined when the ATA is used as an
autocollimator to measure the HRMA line of sight (LOS) is shown
in E445910.  The ATA measures the rotation angle of the normal to
the alignment reference mirror (ARM) about the XRCF Y axis, qY,
and the rotation about the XRCF Z axis, qZ.  The correctly
reckoned signs for the rotations are shown in E445910,
corresponding to the two rotation angles when viewed through the
ATA eyepiece reticle.

3.1.5     SCIENCE INSTRUMENT MODULE FIVE AXIS MOUNTCIENCE
INSTRUMENT MODULE FIVE AXIS MOUNT

The coordinate system for the science instrument module (SIM)
five axis mount (FAM) is TBD.

3.1.6     HRMARMA

The coordinate system for the HRMA is defined in EK-5003-100,
section 3.3.11 and Figure 3.3.11-1.

3.1.7     HRMA X-ray Detection SystemRMA X-ray Detection System

The HXDS coordinate system has its axes parallel (to within its
alignment tolerances) to the corresponding axes of the XRCF
coordinate system.  The origin of the HXDA coordinate system is
located at TBD.

3.2  UNITSNITS

3.2.1     Lengthength

Unless otherwise indicated, all dimensions in this document are
in feet and inches.

3.3  TIME DESIGNATIONIME DESIGNATION

If a time tag is not computer or IRIG generated, it shall be
noted with a suffix of "Z" at the end.  For example, 1PM
Universal Time (UT) would be written, "1300Z".
3.4
SUBSYSTEM REQUIREMENTSUBSYSTEM REQUIREMENTS

3.4.1     MSFC SubsystemsSFC Subsystems

3.4.1.1   X-ray Calibration Facility-ray Calibration Facility

The MSFC XRCF is defined in MSFC-SPEC-1837.

3.4.1.2   X-ray Source System-ray Source System

The requirements for the x-ray source system (XSS) are given in
MSFC-SPEC-2229 and the "Draft Specification for the X-ray
Calibration Facility X-ray Source System."

3.4.1.2.1 Source Locationource Location

The source shall be located between STA 27+05+5.

3.4.1.2.2 Knowledge of Source Locationnowledge of Source Location

Knowledge of the source location shall be better than three
inches (TBR).

3.4.1.3   X-ray Monochrometer-ray Monochrometer

The requirements for the x-ray monochrometer (XM) are given in
MSFC-SPEC-2229.

3.4.1.4   Motion Detection Systemotion Detection System

The motion detection system (MDS) requirements are found in
D17830.

3.4.1.5   Autocollimating Telescope Assemblyutocollimating
Telescope Assembly

The ATA requirements can be found in DR XC02, VETA-I Calibration
Requirements.

3.4.1.6   Master Control Computeraster Control Computer

The Master Control Computer requirements are defined in AXAF
Master Control Computer System Requirements Specification, Final
Draft, December 2, 1993, and documents called out therein.

3.4.2     TRW SubsystemsRW Subsystems

3.4.2.1   Command Telemetry Unit Emulatorommand Telemetry Unit
Emulator

Detailed equipment specifications are found in EQ16-0057.

3.4.2.1.1 Functional Purposeunctional Purpose

The CTUE shall provide a common interface between the MCC, SI and
SIM EGSE and the SIs and ISIM RCTUs, as illustrated in TRW DWG
E445900.

3.4.2.1.2 Timing Signalsiming Signals

3.4.2.1.2.1    Frame Synchronizationrame Synchronization

The CTUE shall provide a science header synchronization pulse.

3.4.2.1.2.2    FPSI ClockPSI Clock

The CTUE shall provide a 1.024 MHz clock signal for FPSI use.

3.4.2.1.2.3    CTUE Clock StabilityTUE Clock Stability

The stability of the CTUE clock shall be better than 1 part in
108.

3.4.2.1.2.4     Facility Time StampFacility Time Stamp

The CTUE shall provide an IRIG B time stamp on each major frame.

3.4.2.1.3 Data Formatsata Formats

3.4.2.1.3.1    Number of Formatsumber of Formats

The CTUE shall provide data in flight telemetry formats per EQ16-
0057.

3.4.2.1.3.2    Display of Format in Useisplay of Format in Use

The CTUE shall display what format is in use.

3.4.2.1.3.3    Dissemination of Dataissemination of Data

The CTUE shall utilize point to point socket connection to
transmit commutated data to the SI and SIM EGSE and the MCC.

3.4.2.1.3.4    Data to the Master Control Computerata to the
Master Control Computer

The data delivered to the MCC shall contain the MCC header.

3.4.2.1.3.5    Data to the SI/SIM EGSEata to the SI/SIM EGSE

The data delivered to the SI/SIM EGSE shall not contain the MCC
header.

3.4.2.1.4 Network Interfaceetwork Interface

The CTUE shall have an ethernet interface to the SI and SIM EGSE
and the MCC

3.4.2.1.5 Interface Protocolnterface Protocol

The interface shall utilize TCP/IP socket protocol.

3.4.2.1.6 Cross Commandsross Commands

A cross command is defined as a syntactically correct command
that is issued fromEGSE console for instrument I to instrument J
(I¹J).

3.4.2.1.6.1    Veto of Cross Commandseto of Cross Commands

The CTUE shall prevent the issuance of cross-commands.

3.4.2.1.6.2    Cross Command Alarmross Command Alarm

The CTUE shall have a visible and audible display in case a cross
command is detected.

3.4.2.1.6.3    Command Inhibit on Detection of Cross
Commandommand Inhibit on Detection of Cross Command

The CTUE shall not issue any further commands in the block when a
cross command is detected.

3.4.2.1.6.4    Manual Restart on Cross Commandanual Restart on
Cross Command

The CTUE shall require a manual restart in the case of detected
cross command.

3.4.2.1.7 CTUE Security TUE Security

3.4.2.1.7.1    Password Accessassword Access

The CTUE shall have password access.

3.4.2.1.7.2    Command Acceptanceommand Acceptance

The CTUE will accept internal CTUE commands only from its
keyboard.

3.4.2.1.8 Proximity to RCTUsroximity to RCTUs

The CTUE shall be capable of being located at least 300 feet from
the target RCTUs.
3.4.3
EKC SubsystemsKC Subsystems

3.4.3.1   HRMA Support StructureRMA Support Structure

3.4.3.1.1 Weighteight

The weight of the HRMA and its Support structure shall be less
than 15,000 pounds.

3.4.3.1.2 Decentration and Initial Alignment Tilt
Toleranceecentration and Initial Alignment Tilt Tolerance

These tolerances are in section 0.

3.4.3.1.3 Remote Tilt Capability emote Tilt Capability

The HRMA support structure shall provide the capability to
remotely tilt the HRMA independently about either the Y or Z axis
through the nominal HRMA nodal point(pitch or yaw).

3.4.3.1.3.1    Remote Tilt Range-Pitchemote Tilt Range-Pitch

The HRMA Support structure shall be able to pitch the HRMA at
least +0.5° about the XRCF Y axis.

3.4.3.1.3.2    Remote Tilt Range-Yawemote Tilt Range-Yaw

The HRMA Support structure shall be able to yaw the HRMA at least
+1.5° about the XRCF Z axis.

3.4.3.1.3.3    Remote Tilt Accuracyemote Tilt Accuracy

The required knowledge of the angular position (radial) of the
HRMA optical axis relative to the facility optical axis (FOA)
shall be ± 10 (TBR) arcsecs up to 25 arcminute and can thereafter
degrade linearly up to 1 arcminute at 1.5°.

3.4.3.1.3.4    Remote Tilt Step Sizeemote Tilt Step Size

The required angular step size of the HRMA optical axis relative
to the centerline of the x-ray beam FOA 1shall be less than 10
arcsecs for both pitch and yaw.

3.4.3.1.3.5    Repeatability of Motionepeatability of Motion
Per 6/8/94 telecon I have told EKC to come back within 0.001 inch
in despace and 0.010 decenter. 5 arcmin in clocking.



3.4.3.1.3.5.1  Tip/Tiltip/Tilt
Per 6/8/94 telecon I have told EKC to come back within 0.001 inch
in despace and 0.010 decenter. 5 arcmin in clocking.
The required accuracy of the repeatability of motion of the HRMA
optical axis relative to the FOA shall be + 20 arcsec for both
pitch and yaw.


3.4.3.1.3.5.2  Despaceespace
Per 6/8/94 telecon I have told EKC to come back within 0.001 inch
in despace and 0.010 decenter. 5 arcmin in clocking.
The required accuracy of the repeatability of motion of the HRMA
optical axis relative to the FOA shall be less than 0.001 inch in
depsace.

3.4.3.1.3.5.3  Decenterecenter
Per 6/8/94 telecon I have told EKC to come back within 0.001 inch
in despace and 0.010 decenter. 5 arcmin in clocking.
The required accuracy of the repeatability of motion of the HRMA
optical axis relative to the FOA shall be less than 0.010 inch in
decenter.

3.4.3.1.3.5.4  Clockinglocking
Per 6/8/94 telecon I have told EKC to come back within 0.001 inch
in despace and 0.010 decenter. 5 arcmin in clocking.
The required accuracy of the repeatability of motion of the HRMA
optical axis relative to the FOA shall be less than 5 arcminutes
in clocking.


3.4.3.1.4 Rotation Angle Axis Locationotation Angle Axis Location

Both the horizontal and vertical HRMA support structure rotation
axes shall be within +/- 1.0 (TBR) inches of the HRMA nodal
point.

3.4.3.1.5 Orientation Reportingrientation Reporting

The position of the HRMA shall be reported to the MCC in a timely
and accurate manner after each change of orientation. (TBR).2

3.4.3.1.6 Repositioning Rateepositioning Rate

The time for the HRMA Support structure to orient the HRMA to any
new position within 30 arcminutes (radial)shall not exceed 3
(three) minutes.

3.4.3.1.7 Stability  tability

The following paragraphs give the requirements for the mechanical
stability of the nominal HRMA node point due to mechanical
vibration and thermal causes in the HSS3.  The ambient
environment inside the instrument chamber is defined as 50±1°F
and a pressure range of 760 to 10-6 torr.

3.4.3.1.7.1    Thermal Stabilityhermal Stability

Thermal induced changes are assumed to have a frequency of less
than 0.1 Hz.  It is assumed that the XRCF is performing per MSFC-
SPEC-1837.4

3.4.3.1.7.1.1  Transverse Stabilityransverse Stability

The HRMA support structure contribution to the Y-Z plane motion
of the HRMA node due to thermal causes shall not exceed 0.000481
inches/axis during any 24 hour time interval.

3.4.3.1.7.1.2  Axial Stabilityxial Stability

The HRMA support structure contribution to motion along the
X-axis of the x-ray detector due to thermal effects shall not
exceed 0.00095 inches during any 24 hour time period.

3.4.3.1.7.2    Vibration Induced Motionibration Induced Motion

Vibration induced motions are assumed to have a frequency of 0.1
Hz or greater.

3.4.3.1.7.2.1  Transverse Stabilityransverse Stability

The HRMA support structure contribution to the Y-Z plane motion
of the HRMA node due to vibration shall not exceed 0.000095
inches/axis.


3.4.3.1.7.2.2  Axial Stabilityxial Stability

The HRMA support structure contribution to motion along the
X-axis of the x-ray detector due to thermal effects shall not
exceed 0.00075 inches during any 24 hour time period.

3.4.3.2   HRMA Forward Contamination CoverRMA Forward
Contamination Cover

The HRMA shall have a forward contamination cover which has the
capability of being remotely and independently opened or closed
over the forward (toward x-ray source) end of the HRMA.

3.4.3.2.1 Obscuration of X-ray Beambscuration of X-ray Beam

When opened, the forward contamination cover shall not obscure or
scatter any of the x-ray radiation being transmitted to the focal
plane for the total HRMA clear aperture.

3.4.3.2.2 Time to Openime to Open

The time required to open the forward contamination cover shall
be no greater than 10 (ten) minutes.

3.4.3.3   HRMA Aft Contamination CoverRMA Aft Contamination Cover

The HRMA shall have an aft contamination cover which has the
capability of being remotely and independently opened or closed
over the aft (away from the x-ray source) end of the HRMA.

3.4.3.3.1 Obscuration of X-ray Beambscuration of X-ray Beam

When opened, the aft contamination cover shall not obscure or
scatter any of the x-ray radiation being transmitted to the focal
plane from the total HRMA clear aperture.

3.4.3.3.2 Time to Openime to Open

The time required to open the aft contamination cover shall be no
greater than 10 (ten) minutes.

3.4.3.4   Purge Capabilityurge Capability

The HRMA internal pressure shall be equal to or greater than the
ambient pressure of the IC during pressurization,
depressurization and ambient pressure phases.

3.4.3.5   OTG MountTG Mount

The OTG mount and insertion/retraction mechanism shall mount to
the HRMA support structure just to the rear of the HRMA
postcollimator.  This structure shall provide the following
functions:
          
          1.   Support the 2 OTGs during x-ray calibration of the
          FPSIs.
          
          2.   Provide for the remote insertion and retraction of
          the OTGs during x-ray test.

3.4.3.5.1 OTG Position Requirements TG Position Requirements

The OTG mount & insertion/retraction mechanism shall be capable
of positioning either of the OTGs into the x-ray beam behind the
HRMA, perpendicular to the HRMA axis, centered on the HRMA axis
with the roll angle of the grating surface about the HRMA optical
axis set to a predetermined value.

3.4.3.5.2  OTGs in UseOTGs in Use

Only one OTG shall be in the x-ray beam at a time.

3.4.3.5.3 Obscuration of X-ray Beam by Retracted OTGbscuration of
X-ray Beam by Retracted OTG

When retracted they shall not obscure the x-ray line of sight
from the HRMA to the FPSIs over the entire HRMA clear aperture.

3.4.3.5.4 OTG Alignment Tolerances TG Alignment Tolerances

OTG static alignment tolerances are given in section 0.

3.4.3.5.4.1    OTG Stability TolerancesTG Stability Tolerances

OTG stability tolerances are given in section 0.

3.4.3.5.4.2    OTG Repeatability TolerancesTG Repeatability
Tolerances

OTG stability tolerances are given in section 0.

3.4.3.5.5 Insertion/Retraction Timensertion/Retraction Time

The OTG mount shall be able to insert or retract either OTG in 3
(three) minutes.

3.4.3.5.6 Position Feedbackosition Feedback

The OTG mount shall provide a means of determining the status
(inserted or retracted) for each OTG.

3.4.3.5.7 OTG Insertion/Retraction Speed and AccelerationTG
Insertion/Retraction Speed and Acceleration

The rate at which the OTGs are inserted and retracted shall not
impart loads to the OTGs that exceed 80% of maximum flight loads.

3.4.3.6   Movement of the Aft Contamination Cover and OTGsovement
of the Aft Contamination Cover and OTGs

The remotely operated control system governing the movement of
the ACC and the OTGs shall permit travel of only one unit at a
time.
3.4.3.7
HRMA Shutter AssemblyRMA Shutter Assembly

The HRMA shutter assembly shall mount to the HRMA support
structure and shall be located between the HRMA and the focal
plane.  This assembly shall provide the following functions:

          
          1.   Shutter the clear x-ray aperture of all but a
          given P/H mirror pair, for each  mirror pair.
          
          2.   Shutter the clear x-ray aperture of that mirror
          pair so that only one of four 90°   8815annular
          quadrant sectors centered on the HRMA is clear and
          transmitting x-ray energy.
          
          3.   The shutter quadrants shall be oriented such that
          they are symmetric with respect to the Y and Z axes.
          
          4.   Each individual shutter blade shall operate
          independently of the open/closed status of any and all
          of the other shutters.


3.4.3.7.1  Shutter Blade ClockingShutter Blade Clocking

The HSA shall be aligned to the facility coordinate system such
that the compound angle between the HSA X axis and the FOA is
less than 1°.  Knowledge of this clocking shall be 5 arcminutes
or less.

3.4.3.7.2  Shutter Blade DesignationShutter Blade Designation

The designations for the shutter blades is defined in DWG
E445909.

3.4.3.7.3 Simultaneity of Shutter Operationimultaneity of Shutter
Operation

Any combination of individual shutter blades shall be capable of
being simultaneously operated.

3.4.3.7.4 Remote Operationemote Operation

The HRMA shutter assembly shall provide the capability to be
remotely operated.

3.4.3.7.5 Time to Reconfigureime to Reconfigure

The time required to reconfigure the HSA shall be no greater than
10 seconds.

3.4.3.7.6 Obscuration of x-ray beambscuration of x-ray beam

When any x-ray shutter is open, it shall not obscure or scatter
any of the x-ray radiation being transmitted to the focal plane
for the desired clear aperture.

3.4.3.7.7 Shutter Opacityhutter Opacity

The HSA blades shall have an opacity to x-rays between 80 eV and
10 keV equal to 0.115 inches of aluminum.

3.4.3.7.8 Position Feedbackosition Feedback

The HSA shall provide a means of determining the status (open or
closed) of the shutter assembly.


3.4.3.7.9 HSA Control InterfaceHSA Control Interface

The HSA shall be controlled from SAO supplied equipment. The
interface is documented in XC05.6

3.4.3.8   1-g Offloader-g Offloader

The 1-g off-loader will shall as a goal7 limit the difference due
to 1g effects between on-orbit and ground encircled energy at any
diameter greater than or equal to 0.010 mm and centered on the
on-axis focal point to 15% (TBR) of the on-orbit value with
uncertainty less than 5% (TBR) of the on-orbit value.

3.4.4          SAO Subsystems SAO Subsystems

3.4.4.1   HRMA X-ray Detector SystemRMA X-ray Detector System

A collection of specialized detectors, the HXDS, shall be
developed and operated for x-ray calibration  rehearsal, HRMA and
HRMA/SI calibration.

3.4.4.1.1 Image Plane Detectorsmage Plane Detectors

Detectors at the HRMA entrance aperture in the instrument chamber
shall include the image plane proportional counters with variable
apertures image plane solid-state detectors with apertures and,
an image plane high speed imager.  Detailed requirements can be
found in MSFC-SPEC-2229.

3.4.4.1.2 Beam Normalization Detectorseam Normalization Detectors

BNDs shall include the following.

3.4.4.1.2.1    BND-500ND-500

Beam normalization detectors shall be required at building 500.
Detailed requirements can be found in MSFC-SPEC-2229.

3.4.4.1.2.2    BND-HND-H

Beam normalization detectors shall be required in the Instrument
Chamber at the HRMA entrance aperture.  Detailed requirements can
be found in MSFC-SPEC-2229.

3.4.5
     BECD Subsystems

3.4.5.1   Five Axis Mountive Axis Mount

a.   provide mechanical interfaces between the ISIM and XRCF
provided detector end test bench

b.   provide a proper thermal interface between the ISIM and XRCF

c.   provide remotely operated and controlled 5 degree of freedom
(DOF) motion control of the ISIM, specifically translation in the
±X, ±Y and ±Z directions and rotation about the Y and Z axes.

d.   provide initial alignment capability in all 6 DOF

e.   provide a mechanical interface for the GFE motion detection
system optical point source

f.   records history of ISIM positioning in the x-ray focal plane

g.   records and controls ISIM temperatures.

h.   provide an MDS interface.

3.4.5.2   Physical Requirementshysical Requirements

3.4.5.3   Envelopenvelope

The FAM will be installed in the instrument chamber and must
remain within the space and volume envelope as defined in DWG
E445700 sheet 2 and  sheet 3, view JJ and KK.  DWG E445700 can be
found in DR XC05, Book 2.

3.4.5.4   Weighteight

The weight of the FAM shall be less than 12,000 pounds.

3.4.5.5        Load Test Load Test

The FAM shall be proof tested per MSFC-STD-126E.

3.4.5.6        Failures  Failures

There shall be no credible failure modes which could propagate
across the interface to the ISIM and cause the failure or loss of
the ISIM.

3.4.5.7   MDS Source AccommodationDS Source Accommodation

3.4.5.7.1      Location of Source  Location of Source

The MDS source shall be located at the following location as
expressed in the XRCF coordinate system:

          X= -373 ± 10 inches
          Y= -4.76 ± 0.050 inches
          Z= -2.75 ± 0.050 inches.
          
3.4.5.7.2 Source Interface Location Stability Relative to SI Aim
Pointource Interface Location Stability Relative to SI Aim Point
          
The interface for the MDS source shall maintain the source
relative position with respect to a given SI aim point to within
±50 µm in the X direction and 1.50 µm radial in the transverse
(Y-Z) plane, as a goal, for periods of up to 2 hours.

3.4.5.8   External Thermal Interfacexternal Thermal Interface

The Cryoshround Assembly (CSA) shall be designed so as to not
prevent the establishment of the a thermal condition of 50° ± 2°F
within the instrument chamber.

3.4.5.9   Internal Thermal Interfacenternal Thermal Interface

The CSA shall simulate the space thermal environment such that
the heat generated in the SIM is adequately removed.

3.4.5.10  Operating Modesperating Modes

The operating mode shall be selectable from the FAM control
console.

3.4.5.11  Static Modetatic Mode

In the static mode, the FAM is commanded to a fixed position.
The FAM remains stationary subject to the relevant specifications
contained herein.

3.4.5.12  Motionotion

The FAM must be able to translate the SIM with the following
characteristics:

3.4.5.12.1     X-axis:  Range-axis:  Range

The  X- axis range shall be at least ±1.0 inches.  The range of
motion is to be centered about the HRMA finite conjugate focal
point.

3.4.6     Initial Alignment to Facility Optical Axisnitial
Alignment to Facility Optical Axis

The ISIM shall be aligned to the facility optical axis (FOA) to
the tolerances given in the following paragraphs.

3.4.6.1   Initial Alignment in Xnitial Alignment in X

The FAM shall place the ISIM within 0.100 inches of the nominal
focal point.

3.4.6.2   Initial Lateral Alignmentnitial Lateral Alignment

The FAM shall place the ISIM such that the initial focal plane
aim point lateral offset (Y-Z plane) is within 0.100 inches
(radial) of the nominal focal point of the HRMA.

3.4.6.3   Initial Alignment of Normal to SIM Focal Planenitial
Alignment of Normal to SIM Focal Plane

The FAM shall align the normal of the SIM focal plane to within
120 arcseconds (radial) of the FOA.

3.4.6.4   Initial Rotation about FOAnitial Rotation about FOA

The FAM shall align the ISIM such that the line containing the
nominal SI aim points forms an angle of  90°±0.25° with respect
to the XRCF Y axis.

3.4.6.5   X-axis: Mechanism Resolution-axis: Mechanism Resolution

The X-axis step size shall be no larger than 0.0005 inches.

3.4.6.6   X-axis: Position Sensor Resolution-axis: Position
Sensor Resolution

The X-axis step resolution shall be no larger than 0.0005 inches.

3.4.6.7   X-axis:  Rate-axis:  Rate

The rate of translation in the X-axis shall be at least 0.1
inch/minute.

3.4.6.8   Y-axis:  Range -axis:  Range

The Y-axis range shall be at least ±7 inches. The range of motion
is to be centered about the HRMA finite conjugate focal point.

3.4.6.9   Y-axis:  Step Size-axis:  Step Size

The Y-axis step size shall be no larger than 0.005 inches.

3.4.6.10  Y-axis:  Resolution-axis:  Resolution

The Y-axis step size resolution shall be no larger than 0.005
inches.

3.4.6.11  Y-axis:  Rate-axis:  Rate

The rate of translation in the Y-axis shall be at least 1
inch/minute.

3.4.6.12  Z-axis:  Range-axis:  Range

The range of motion shall be arranged such that each SI can be
located ±7 inches (±1°) from the nominal HRMA finite conjugate
focal point in the Z direction.

3.4.6.13  Z-axis:  Step Size-axis:  Step Size

The Z-axis step size shall be no larger than 0.005 inches.

3.4.6.14  Z-axis:  Resolution-axis:  Resolution

The Z-axis step size resolution shall be no larger than 0.005
inches.

3.4.6.15  Z-axis:  Rate-axis:  Rate

The rate of translation in the Z-axis shall be at least 1
inch/minute.

3.4.6.16  Dither  Modeither  Mode

The ISIM FAM dither mode is intended to be a simulation of the
spacecraft dither mode.  In this operating mode the image is
spread accross the focal place by a series of small movements.
In the XRCF this will be accomplished by driving the ISIM to the
desired location and invoking a dither program.  This will move
the ISIM in small steps dwelling at each location for a
commandable periods of time, before slewing to the next position.

3.4.6.16.1     Step Sizetep Size

The step size, d, for the dither mode shall be 11±1 mm
(0.433±0.040 mil)

3.4.6.16.2     Range of Capabilityange of Capability

The dither mode shall be capable of operation  at any location
over the full FAM range.

3.4.6.16.3     Positonal Accuracy-Absoluteositonal Accuracy-
Absolute

When commanded, the SIM shall go to the commanded position ±1
step.

3.4.6.16.4     Postional Accuracy-Relative to Starting
Positionostional Accuracy-Relative to Starting Position

The relative accuracy  (per axis) in the knowlegde of the
posistion of the SIM shall be ±2mm for any 2 mm sub-range.

3.4.6.16.5     Dwell Time

The dwell time at each location shall be specified in a data file
in seconds supplied by the FPSI teams prior to x-ray calibration.

3.4.6.16.6     Motion Controlotion Control

The FAM dither mode shall be capable of commanding the FAM to
begin from a location, move to another and dwell for a specified
period of time.  The FAM control software shall receive data from
an FPSI supplied data file.  The data file shall contain a list
of ASCII values giving the Y and Z coordinates relative to the
current location, and the dwell time in seconds at each location.

3.4.6.16.7     Data ata

3.4.6.16.7.1   Data Displayata Display

The SIM control console shall display the position of the SIM in
near real time.

3.4.6.16.7.2   Data Transmissionata Transmission

At the conclusion of a dither measurement sequence, the data
history of postions, dwell and slew times and all other relevant
data shall be IRIG time stamped and archived to the MCC.
3.4.7
Stability  tability

The following paragraphs give the requirements for the mechanical
stability of the nominal SIM aim point due to mechanical
vibration and thermal causes.  The ambient environment inside the
instrument chamber is defined as 50±1°F and a pressure range of
760 to 10-6 torr.

3.4.7.1   Thermal Stabilityhermal Stability

Thermal induced changes are assumed to have a frequency of less
than 0.1 Hz.

3.4.7.1.1 Transverse Stabilityransverse Stability

The FAM contribution to the Y-Z plane motion of the x-ray
detector due to thermal causes shall not exceed 0.000481
inches/axis during any 24 hour time interval.

3.4.7.1.2 Axial Stabilityxial Stability

The FAM contribution to motion along the X-axis of the x-ray
detector due to thermal effects shall not exceed 0.00095 inches
during any 24 hour time period.

3.4.7.2   Vibration Induced Motionibration Induced Motion

Vibration induced motions are assumed to have a frequency of 0.1
Hz or greater.

3.4.7.2.1 Transverse Stabilityransverse Stability

The FAM contribution to the Y-Z plane motion of the x-ray
detector due to vibration shall not exceed 0.000095 inches/axis.


3.4.7.2.2 Axial Stabilityxial Stability

The FAM contribution to motion along the X-axis of the x-ray
detector due to vibration induced motions  shall not exceed
0.00075 inches during any 24 hour time period.



3.4.8     Rotation

The FAM must be able to rotate the SIM with the following
characteristics:

3.4.8.1   X-axis: Rotation-axis: Rotation

Rotation of about the X-axis is not required.

3.4.8.2   Y-axis:  Range-axis:  Range

The range of rotation about the Y- axis shall be at least ±1°.

3.4.8.3   Y-axis:  Step Size-axis:  Step Size

The step size in rotation about the Y-axis shall be no greater
than 10 arcsecs.

3.4.8.4   Y-axis:  Resolution-axis:  Resolution

The resolution of angular orientation in rotation about the Y
axis shall be no larger than 10 arcsecs.

3.4.8.5   Y-axis:  Rate-axis:  Rate

At least  0.1 degree/minute.

3.4.8.6   Z-axis:  Range-axis:  Range

The range of rotation about the Z- axis shall be at least ±1°.

3.4.8.7   Z-axis:  Step Size-axis:  Step Size

The step size in rotation about the Y-axis shall be no greater
than 10 arcsecs.

3.4.8.8   Z-axis:  Resolution-axis:  Resolution

The resolution of angular orientation in rotation about the Y
axis shall be no larger than 10 arcsecs.

3.4.9     FAM Controller SoftwareAM Controller Software

3.4.9.1   Communicationommunication

The FAM controller must be able to communicate with the MSFC
supplied Master Control Computer. (MCC)  The MCC issues messages
to the target system via ethernet to queue test activity.  Upon
receipt of the message, the operator of the FAM will respond
appropriately. The FAM controller should also reply to the
messages as tasks are completed.   Thus, it is recommended that
the operating system for the FAM controller be a multitasking
windows capable system.

3.4.9.2   Data Recordingata Recording

A full history of positional sensor and thermal data shall be
maintained.  These records will be transferred to the MCC, which
is also on the LAN.

3.4.10    ACIS Cryogenic InterfaceCIS Cryogenic Interface

The SIM shall accommodate the ACIS cryogenic interface.

3.4.11    Accessccess

The SIM shall accomodate access to the ACIS high speed data
connector.

3.4.12    Vacuum Interfaceacuum Interface

The SIM shall provide an interface for vacuum capability.
3.4.13
HRC SubsystemsRC Subsystems

3.4.13.1  Electrical Ground Support Equipment-HRClectrical Ground
Support Equipment-HRC

3.4.13.1.1     Command Validationommand Validation

The HRC electrical GSE (EGSE) shall validate all commands sent to
the HRC.

3.4.13.1.2     Telemetry Decommutationelemetry Decommutation

The HRC EGSE shall be able to decommutate any and all data
necessary to operation of the HRC that is transmitted to it by
the CTUE.

3.4.13.2  Verificationerification

The HRC is responsible for critical command verification.

3.4.13.3  Command Formatommand Format

The HRC shall be compatible with the command format given in
paragraph 3.2.1.2.3 of EQ16-0057.

3.4.13.4  SI Command LogI Command Log

The HRC shall maintain a log of SI commands.

3.4.14    ACIS SubsystemsCIS Subsystems

3.4.14.1  Electrical Ground Support Equipment-ACISlectrical
Ground Support Equipment-ACIS

3.4.14.1.1     Command Validationommand Validation

The ACIS electrical GSE (EGSE) shall validate all commands sent
to the ACIS.

3.4.14.1.2     Telemetry Decommutationelemetry Decommutation

The ACIS EGSE shall be able to decommutate any and all data
necessary to operation of the HRC that is transmitted to it by
the CTUE.

3.4.14.2  Verificationerification

The ACIS is responsible for critical command verification.

3.4.14.3  Command Formatommand Format

The ACIS shall be compatible with the command format given in
paragraph 3.2.1.2.3 of EQ16-0057.

3.4.14.4  SI Command LogI Command Log

The ACIS shall maintain a log of SI commands.

3.4.15    LETG SubsystemsETG Subsystems

TBD.

3.4.16    HETG SubsystemsETG Subsystems

TBD.

3.4.17    ASC SubsystemsSC Subsystems

TBD.

3.4.18
Other Systemsther Systems

3.4.18.1  X-ray  Rehearsal Optic-ray  Rehearsal Optic

The x-ray rehearsal optic (XSO) is to be used during the
rehearsal period. It purposes are to simulate the HRMA in such a
way that the test equipment, procedures and personnel my be
properly wrung out to facilitate an efficient operation during
HRMA and HRMA/SI calibration.

3.4.18.1.1     Spot Sizepot Size

The spot size (FWHM) of  the XSO shall be between 25-40 mm.

3.4.18.1.2     Effective Areaffective Area

The effective area of the XSO shall be 20 cm2 or greater.

3.4.18.1.3     f/number/number

The f/number of the XSO shall be bewteen 10 and 16.

3.4.18.2  X-ray  Rehearsal Optic Mount-ray  Rehearsal Optic Mount


3.4.18.2.1     Postitioningostitioning

The XSO mount shall be capable of  locating the XSO as defined in
0.

3.4.18.2.2     Stabilitytability

The XSO mount shall meet the stability requirements as called out
in 0.

4    ALIGNMENT/STABILITY REQUIREMENTS AND
ALLOCATIONSLIGNMENT/STABILITY REQUIREMENTS AND ALLOCATIONS


4.1  HRMA TO FACILITY OPTICAL AXISRMA TO FACILITY OPTICAL AXIS

The HRMA support structure shall provide the capability to
initially align the HRMA optical axis coincidentally with the FOA
prior to x-ray testing.  The allowable initial alignment error
tolerances relative to the centerline of the x-ray beam FOA are
given below.
8
4.1.1     AXIAL LOCATIONXIAL LOCATION
Per telecon on 6/8/94 with Vallimont and Johnston value changed
to +/- 0.5 inch.
The axial location of the HRMA CAP mid-plane is at XXRCF = 0 ±
0.5 inch.

4.1.2     LATERAL DECENTRATION OF HRMA NODAL POINT:ATERAL
DECENTRATION OF HRMA NODAL POINT:

The HRMA support structure shall be capable of locating the HRMA
nodal point to within 0.1 inch (radial) of the facility optical
axis.

4.1.3     TILT OF THE HRMA OPTICAL AXIS:ILT OF THE HRMA OPTICAL
AXIS:

The HRMA shall be aligned such that the angle between the HRMA
optical axis and the facility optical axis is 15 arcsec (radial).

4.1.4     HRMA CLOCKINGRMA CLOCKING
Per telecon on 6/8/94 with Vallimont and Johnston value changed
to +/- 0.5 inch.
The HRMA shall be aligned such that the clocking angle about the
FOA is less than 32.5 arcminutes.

4.2  HRMA TO OTGRMA TO OTG

All alignment tolerances between the HRMA and OTGs are 3 sigma
tolerances.

4.2.1     Static Alignment Tolerancestatic Alignment Tolerances

The grating alignment references are a set of features located on
the OTGs.  When the term GAR is used in the sections below it is
taken to mean the appropriate reference as called out on the OTG
interface control drawing, TRW DWG 301331. 9

4.2.1.1   Static Alignment in Xtatic Alignment in X

The GRIM shall locate the grating alignment reference (GAR) to ±
0.040 inch of its desired location in X.

4.2.1.2   Static Decentertatic Decenter

The GRIM shall locate the grating to within ±0.0174 inch of its
desired location in Y,Z plane.

4.2.1.3   Static Alignment in Rotation About Xtatic Alignment in
Rotation About X

The GRIM shall locate the GAR to within ±3.75 arcmin of its
desired location in rotation about X.

4.2.1.4   Static Alignment in Rotation About Y Tip/Tilttatic
Alignment in Rotation About Y

The GRIM shall locate the GAR to within ±2 arcmin of its desired
location in rotation about Y. The GRIM shall orient the GAR such
that the angle between the GAR-X axis and the FOA is less than
3.54 arcminutes.10

4.2.1.5   Static Alignment in Rotation About Ztatic Alignment in
Rotation About Z

The GRIM shall locate the GAR to within ±2 arcmin of its desired
location in rotation about Z./

4.2.2     Stability Tolerancestability Tolerances

4.2.2.1   Stability in Xtability in X

The GRIM shall maintain the GAR stable in the axial (X) direction
with respect to its intial axial location upon insertion to
within ± 0.0118 inch11.

4.2.2.2   Decenter Stabilityecenter Stability

The GRIM shall maintain the GAR stable in decenter (radial motion
in the Y,Z plane)  with respect to its intial  location upon
insertion to within ±0.0052 inch12.

4.2.2.3   Stability in Rotation About Xtability in Rotation About
X

The GRIM shall maintain the GAR stable in rotation about the X
direction with respect to its orientation upon insertion to
within ±1.125 arcminutes.

4.2.2.4   Stability in Rotation About YTip/Tilt13tability in
Rotation About Y

The GRIM shall maintain the GAR stable in rotation about the Y
direction with respect to its orientation upon insertion to
within ±0.5625 arcminutes.The GRIM shall maintain the orientation
to less than 1.06 arcminutes in tip/tilt.

4.2.2.4   Stability in Rotation About Ztability in Rotation About
Z

The GRIM shall maintain the GAR stable in rotation about the Z
direction with respect to its orientation upon insertion to
within ±0.5625 arcminutes.

4.2.3     Repeatability Tolerancesepeatability Tolerances

4.2.3.1   Repeatability in Despaceepeatability in Despace

The GRIM mechanism shall place the GAR in the axial (X) direction
with respect to its previous axial location upon insertion to
within ±0.0118 inch.

4.2.3.2   Repeatability in Decenterepeatability in Decenter

The GRIM mechanism shall place the GAR in the axial (X) direction
with respect to its previous axial location upon insertion to
within ±0.0052 inch14.

4.2.3.3   Repeatability in Rotation About Xepeatability in
Rotation About X

The GRIM mechanism shall place the GAR in rotation about the X
direction with respect to its orientation upon  the previous
insertion to within ±1.125 arcminutes.

4.2.3.4   Repeatability in Rotation About Y Tip/Tilt15epeatability
in Rotation About Y

The GRIM mechanism shall place the GAR in rotation about the Y
direction with respect to its orientation upon  the previous
insertion to within ±0.5625 arcminutes. The GRIM mechanism shall
have a repeatability of less than 1.06 arcminutes in tip/tilt.

4.2.3.5   Repeatability in Rotation About Zepeatability in
Rotation About Z

The GRIM mechanism shall place the GAR in rotation about the Z
direction with respect to its orientation upon  the previous
insertion to within ±0.5625 arcminutes.

4.3  ENVIRONMENTNVIRONMENT

4.3.1     Stray Lighttray Light

The ambient stray light level within the IC at the x-ray detector
aperture shall be consistent with the stray light levels as given
in MSFC-SPEC-1836, paragraph 4.2.3.1.4.  (NB: these levels apply
only when ACIS is in operation.)

4.3.2      Microseismic Vibration Fundamental
Frequencyicroseismic Vibration

The power spectral density of typical XRCF disturbances is given
in DR XCO5. All GSE shall have a fundamental frequency in excess
of 4 Hz.16

4.3.3     Vacuumacuum

The vacuum environment of the IC shall meet the requirements
given in MSFC-SPEC-1837.

4.3.4     Thermalhermal

The thermal environment of the IC shall meet the requirements
given in MSFC-SPEC-1837.
5
     WORKMANSHIP STANDARDS    WORKMANSHIP STANDARDS

The requirements contained in this section apply to all hardware
that will be resident in the instrument chamber.

5.1  BAKEOUTAKEOUT

All IC resident hardware shall be baked to remove volatiles.
This bakeout shall be in accordance with MSFC-SPEC-1238.  This
"bake out" must occur prior to installation of the  HRMA in  the
XRCF vacuum chamber.

5.1.1     Surface Cleanliness Levelsurface Cleanliness Levels

All IC resident hardware shall be cleaned to a level of 350Å in
accordance with MIL-STD-1246.



5.2  TRAPPED VOLUMESRAPPED VOLUMES

All IC resident hardware shall have no trapped volumes that
preclude the achievement of an IC pressure level of 1x10-6 torr
in 24 hours.
6
     GLOSSARY OF TERMS   GLOSSARY OF TERMS


Aimpoint:                     Any point on either active focal
plane of either FPSI.

Resolution, Mechanism:        The minimum controllable motion
interval that the system is capable of producing.

Resolution, Sensor:           The minimum motion interval that
the sensor is capable of reporting.

Repeatability, Mechanisms,    The extent to which successive
attempts to move to a specific
Uni-directional:              location from a single direction
vary in position.

Repeatability, Mechanisms,    The extent to which successive
attempts to move to a specific
Bi-directional:               location from opposite directions
vary in position.

Repeatability, Sensor:        The variation in reports by a
sensor system produced by successive movements to a specific
location.

Stability, Mechanical:        The ability of a system to maintain
relative position of datums over a given range of environments
with the passage of time.

Stability, Sensors:           The  ability of a sensor system to
report the same measurement when measuring an unchanged quantity
over a given range of environments with the passage of time.

Accuracy, Mechanism:          The degree to which displacements
executed by a positioning system match agreed upon standards.

Accuracy, Sensor:             The degree to which displacements
measured by a system match agreed upon standards.
7
ACRONYMSCRONYMS



ACIS
     AXAF CCD Imaging Spectrometer                 26, 27, 29, 37
ACS
     Architectural Coordinate System                            4
ARM
     Alignment Reference Mirror                                 4
ATA
     Alignment Telescope Assembly                            4, 8
BECD
     Ball Electro-Optics and Cryogenics Division               20
BND
     Beam Normalization Detector                               19
CAP
     Center Aperture Plate                                     34
CTUE
     CTU Emulator        9-11, 28, 29
DWG
     Drawing             4, 9, 17, 20
EGSE
     Electrical Ground Support Equipment            9, 10, 28, 29
EKC
     Eastman Kodak Company                                     12
FAM
     Five Axis Mount     5, 20-26
FOA
     Facility Optical Axis                 12, 13, 17, 21, 22, 34
FPSI
     Focal Plane Science Instrument                     9, 24, 40
FWHM
     Full Width Half Maximum                                   33
GFE
     Government Furnished Equipment                            20
GSE
     Ground Support Equipment                        1, 2, 28, 29
GT
     Guide Tube          4
HETG
     High Energy Transmission Grating                          31
HRC
     High Resolution Camera                                28, 29
HRMA
     High Resolution Mirror Assembly   1, 2, 4, 5, 12-15, 17, 19,     21-23, 33-35,
38
HXDA
     HRMA X-ray Detector Assembly                               5
HXDS
     HRMA X-ray Detection System                            5, 19
IC
     Instrument Chamber  15, 37-39
ICVS
     Instrument Chamber Vacuum System                          10
IRIG
     Intra Range Instrumentation Group                   7, 9, 24
ISIM
     Intergrated Science Instrument Module               9, 20-23
LETG
     Low Energy Transmission Grating                           30
LOS
     Line of Sight       4
MCC
     Master Control Computer                    9, 10, 13, 24, 26
MDS
     Motion Detection System                      2, 4, 8, 20, 21
MSFC
     Marshall Space Flight Center     1, 2, 8, 19, 20, 26, 37, 38
OPS
     Optical Point Source                                       4
OTG
     Optical Transmission Grating                      15, 16, 35
P/H
     Paraboloid/Hyperboloid                                    17
SAO
     Smithsonian Astrophysical Observatory                     19
SI
     Science Instrument  1, 2, 9, 10, 19, 21-23, 28, 29, 33
SIM
     Science Instrument Module                5, 9, 10, 21-25, 27
STA
     Station             4, 8
TBD
     To Be Determined    5, 30-32
TBR
     To Be Resolved      8, 12, 13, 18
TRW
     TRW Incorporated    1, 2, 4, 9
XRCF
     X-ray Calibration Facility         1-5, 8, 12, 20-23, 37, 38
XSO
     X-ray Surrogate Optic                                     33

_______________________________
1XRCCOM-14
2Remove TBR, this is afterall a CDR version of requirements of
hardware that has been through CDA.
3XRCCOM-11
4XRCCOM-12
5XRCCOM-13
6XRCCOM-26
7 XRCCOM-10
8XRCCOM-14
9XRCCOM-16, 49 and 50 Proper text and reference to OTG ICD.
10XRCCOM-17 combine  rotations and equate with flight
requirements.
11Missing units.
12Missing units.
13XRCCOM-17
14Missing units.
15XRCCOM-17
16XRCCOM-15This replacement of fundamental frequency for GSE for
definition of vibration spectrum was agreed to by all parties
involved as the simplest surest way to insure vibrational
stability at XRCF.


Back to the XTC Home Page