Table 1: Calibration Instrument Requirements Summary

Calibration Test

VETA-I Test

AXAF Preflight Calibration

Accuracy

Requirement

Detectors

used

Monitors used

Accuracy

Requirement

Additional Instrument Required

Additional Monitors required

Type

Reason

Type

Reason

FWHM

0.05 arcsec

1 small flow, proportional counter pinhole scanner.

1 small flow, 1 small sealed

proportional counters @

VETA-I

same as VETA-I

.....................................................

Encircled Energy

2%, 5 E's
0.27 - 2 keV

1%, 30 E's, 0.1-10 keV

SSS, stable, fast x-ray source

high-energy

continuum subtration, speed, accuracy

~3 more medium flow proportional counters. SSS

beam spatial, temporal & spectral uniformity

Mirror

Effective

Area

5%, 5 E's
0.27 - 2 keV

1%, 30 E's, 0.1-10 keV

1% E/ E = 103, 0.1-10 keV

Synchrotron, mono., reflectometer

monochromatic beam

(small flow proportional counters, pin diodes)

beam temporal uniformity

Grating

efficiency

.....................................................

2%,E/ E = 2000, 0.1-10 keV

HRMA PRF details
(A/A)

.....................................................

inner: 0.5%, intermed.: 2%

HSI, small x-ray spot

speed: 2 yrs -> 5 mo.

(~3 more medium flow proportional counters)

beam spatial & temporal uniformity

.....................................................

MONO: 0.1-10

eliminate continuum

.....................................................

outer: 15% over 30 mm field, (A/A)

HRC

speed (solid angle)

.....................................................

MONO

eliminate continuum

Grating

Energy

Resolution
( FWHM/FWHM)

.....................................................

LETG: 2%,.095-7 keV

Penning source

narrow 0.095 keV line

.....................................................

HEG: 2%

SSS




.....................................................

MEG: 2%,0.4-7 keV

MONO:0.4-1.0

narrow 0.4 keV line



Grating PRF wings

.....................................................

1%
(of peak)

SSS

high-energy

continuum subtraction

SSS

beam spectral uniformity

HRMA ghost images

.....................................................

.......................................

HRC

rapid detection @ high sensitivity

3 more flow ...

spatial, spectral

Structure

fluorescence

.....................................................

look for features from deposits, obstructions

(Flight gratings)

will show up as spectral features






Back to Introduction