To: Distribution

From: M. Zombeck

Subject: Calibration of Neat translation stages.

Date: 25 February 1999

The linearity and repeatability of three 8 inch and one 18 inch HRC Laboratory NEAT translation stages were determined using an HP 5529A Dynamic Calibrator (laser interferometer).

A summary of the results is given below. The individual stage calibration results are provided in postscript and can be obtained by clicking on the appropriate table entry.


DescriptionSerial No. Leadscrew Cumulative Error(1) Lead Screw Periodic Error(2) Repeatability(3)Home Position Repeatability Postscript File
18" HM-180000125 microns 15 microns < 1 micron 4 micronssn001.ps
8" LM-80000212 2* < 1 6 sn002.ps
8" LM-800003< 1 7 < 1 4 sn003.ps
8" LM-80000415 2 < 1 1 sn004.ps
Specs. (4)
-
20 (46) (5) 10 < 2.5 < 2.5 --------

(1) Runout for total travel

(2) Peak-to-peak; 2 mm period

(3) Neat recommended repeatability test (run open loop):

Move stage to mid-stage position and reset interferometer to 0.0000

Move +5 mm and take reading

Move -5 mm and take reading

Move -5 mm and take reading

Move +5 mm and take reading

Repeat the above 3 times

* First 130 mm

(4) Specifications are for "precision" stages; we purchased "commercial".

(5) Value for 18" stage