To: Distribution
From: M. Zombeck
Subject: Calibration of Neat translation stages.
Date: 25 February 1999
The linearity and repeatability of three 8 inch and
one 18 inch HRC Laboratory NEAT translation stages were determined using an
HP 5529A Dynamic Calibrator (laser interferometer).
A summary of the results is given below. The individual
stage calibration results are provided in postscript and can
be obtained by clicking on the appropriate table entry.
| Description | Serial No. | Leadscrew Cumulative Error(1) | Lead Screw Periodic Error(2) | Repeatability(3) | Home Position Repeatability | Postscript File |
| 18" HM-1800 | 001 | 25 microns | 15 microns | < 1 micron | 4 microns | sn001.ps |
| 8" LM-800 | 002 | 12 | 2* | < 1 | 6 | sn002.ps |
| 8" LM-800 | 003 | < 1 | 7 | < 1 | 4 | sn003.ps |
| 8" LM-800 | 004 | 15 | 2 | < 1 | 1 | sn004.ps |
| Specs. (4) | 20 (46) (5) | 10 | < 2.5 | < 2.5 | -------- |
(1) Runout for total travel
(2) Peak-to-peak; 2 mm period
(3) Neat recommended repeatability
test (run open loop):
Move stage to mid-stage position and reset interferometer to 0.0000
Move +5 mm and take reading
Move -5 mm and take reading
Move -5 mm and take reading
Move +5 mm and take reading
Repeat the above 3 times
* First 130 mm
(4) Specifications are for "precision" stages; we purchased "commercial".
(5) Value for 18" stage